Prompt Inoculation System-D Inoculation Wand (2019-12-16)
- Starting date:
- December 16, 2019
- Posting date:
- July 3, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73463
Last updated: 2020-07-03
Affected Products
Prompt Inoculation System-D Inoculation Wand
Reason
As a result of customer reports and an internal investigation, Beckman Coulter concluded that the Prompt Inoculation System-D Lot Number 20201103T contains boxes of incorrect Prompt wands. The Prompt Inoculation System-D wands are used for MicroScan panel inoculum preparation. The supplier manufactures a similar wand for another company, and they placed the incorrect wands in the MicroScan Prompt System box.
Affected products
Prompt Inoculation System-D Inoculation Wand
Lot or serial number
20201103T
Model or catalog number
B1026-10D
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES