Health product recall

Proteus XR/a General Purpose Radiographic System

Last updated

Summary

Product
Proteus XR/a General Purpose Radiographic System
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Proteus XR/a General Purpose Radiographic System

More than 10 numbers, contact manufacturer.

2259988
2268970
2259973-3

Issue

GE Healthcare has become aware that steel cables in the wall stand for certain Proteus XR/a systems serviced by GE Healthcare have not been replaced at the intervals detailed in the preventative maintenance action list (GEHC service manual 2273022-100).

Recall Start Date: November 4, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Ge Hualun Medical Systems Co. Ltd.

No.1 Yong Chang North Road, Beijing, Beijing, China, 100176

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76599

Get notified

Receive notifications for new and updated recalls and alerts by category.