Health product recall

QinFlow Warrior Lite Base Unit

Last updated

Summary

Product
QinFlow Warrior Lite Base Unit
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.  

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

Warrior Lite Base Unit

More than 10 numbers, contact manufacturer.

QIF-03

Issue

The warrior lite base unit operates in conjunction with a battery and a disposable unit to warm blood and IV fluids prior to administration to patients.
Following a technical root-cause investigation into field complaints involving thermal deformation of disposable units, quality in flow determined that an internal mechanical shield was incorrectly assembled on the Printed Circuit Board Assembly (PCBA) during manufacturing. Under harsh environmental conditions or significant mechanical impact, an incorrectly assembled shield may detach from the PCBA and cause an electrical short circuit.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Quality In Flow (QinFlow) Ltd.

23 Ha-Yetsira St., Petah Tikva, Israel, 4951251

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-83001