SCS SCREWDRIVER FOR RATCHET (2021-05-26)
- Starting date:
- May 26, 2021
- Posting date:
- June 3, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75725
Last updated:
2021-06-03
Affected Products
SCS SCREWDRIVER FOR RATCHET
Reason
The tip of the screwdrivers does not have correct profile therefore does not properly fit the screw head and cannot pick up screws. No patient harm has occurred, however this can result in reduction in aspiration protection.
Affected products
SCS SCREWDRIVER FOR RATCHET
Lot or serial number
- EHG85
- FAP78
Model or catalog number
- 46.401
- 46.402
Companies
- Manufacturer
-
Institut Straumann AG,
Peter Merian-Weg 12,
Basel
4002
SWITZERLAND