Health product recall

ShockPulse Lithotripsy Transducer

Brand(s)
Last updated

Summary

Product
ShockPulse Lithotripsy Transducer
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
ShockPulse Lithotripsy Transducer More than 10 numbers, contact manufacturer. SPL-T

Issue

Olympus found that the SPL-T transducer may fail to start. In addition, the body of the transducer handpiece may gradually increase in temperature during use, potentially before its expected lifespan. users are advised to have a backup transducer and sterilized probe ready prior to beginning a procedure.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Gyrus ACMI, Inc

9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78383

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe