SICAT Implant V2.0 (2020-07-14)
- Starting date:
- July 14, 2020
- Posting date:
- August 14, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73683
Last updated:
2020-08-14
Affected Products
SICAT Implant V2.0
Reason
SICAT IMPLANT is a standalone software application, i.e. a radiological visualization software for diagnosis and dental implant planning.
We found that implant planning data exported from SICAT IMPLANT are not correct for one specific export format.
The error in the export format might cause dentists to manufacture (or have manufactured) surgical guides which are incorrect.
Using those guides might cause drilling into the jaw to be either not deep enough or to deep.
A dentist might have to stop a surgical procedure, or the drilling when to deep might cause damage to structures within the jaw.
Affected products
SICAT Implant V2.0
Lot or serial number
Sicat Implant V2.0 software
Model or catalog number
SICAT IMPLANT
Companies
- Manufacturer
-
SICAT GmbH & Co. KG.
BRUNNENALLEE 6
BONN
53177
GERMANY