SIGNA™ Artist and SIGNA™ Artist Evo
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
SIGNA™ Artist |
PGRART2500010 |
SIGNA ARTIST |
SIGNA™ Artist Evo |
IPGEV2400062SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEVO2500008 |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2400132SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2500201SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2400103SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2400090SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2400110SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2500222SC |
00195278481382 |
SIGNA™ Artist Evo |
IPGEV2400083SC |
00195278481382 |
Issue
GE Healthcare has become aware of an exposed sharp edge inside the SIGNA™ Artist Evo and SIGNA™ Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity.