Health product recall

SIGNA™ Explorer

Last updated

Summary

Product
SIGNA™ Explorer
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

SIGNA™ Explorer

780964EXP
604822MRTS
418691MR

6433266-1

Issue

GE Healthcare became aware that for certain MR systems a scan can be resumed following a power monitor trip when the '6-minute average sar' is above the limit indicated on the 'sar display.'

Recall start date: February 12, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Ge Healthcare (Tianjin) Company Limited

No. 266 Jingsan Road, Tianjin, China, 300308

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76989

Get notified

Receive notifications for new and updated recalls and alerts by category.