Health product recall

SmartPort Implantable Ports

Last updated

Summary

Product
SmartPort Implantable Ports
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information.  

Affected products

Affected products

Lot or serial number

Model or catalogue number

SmartPort Implantable Ports

A2926044

H787CT80LPPDVI0

SmartPort Implantable Ports

A2926056

H787CT60LPPDVI0

Issue

AngioDynamics, Inc., is conducting a voluntary medical device field safety corrective action or recall to the end-user level based on the awareness that components within specific models and lots of the SmartPort Plastic Implantable Ports or Smart Port CT port kits may not meet dimensional specifications and will not allow for a secure connection between the port chamber and the catheter. The affected SmartPort Plastic Implantable Ports being recalled (removed) have the potential to contain an incorrect sized introducer/sheath, locking collar, or port. The SmartPort Plastic Implantable Port kits may be packaged with an incorrect sized introducer sheath that will not allow advancement of the catheter through the peelable sheath. The SmartPort Plastic and Smart Port CT kits may also be packaged with incorrect sized locking collar or port assemblies which may not allow a secure connection of the locking collar to the port stem or not allow the catheter to pull off the port stem once assembled with a smaller locking collar.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Angiodynamics, Inc. Also D.B.A. As Navilyst Medical , Inc.

26 Forest Street, Marlborough, Massachusetts, United States, 01752

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-83002

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe