SmartPort Implantable Ports
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
SmartPort Implantable Ports |
A2926044 |
H787CT80LPPDVI0 |
SmartPort Implantable Ports |
A2926056 |
H787CT60LPPDVI0 |
Issue
AngioDynamics, Inc., is conducting a voluntary medical device field safety corrective action or recall to the end-user level based on the awareness that components within specific models and lots of the SmartPort Plastic Implantable Ports or Smart Port CT port kits may not meet dimensional specifications and will not allow for a secure connection between the port chamber and the catheter. The affected SmartPort Plastic Implantable Ports being recalled (removed) have the potential to contain an incorrect sized introducer/sheath, locking collar, or port. The SmartPort Plastic Implantable Port kits may be packaged with an incorrect sized introducer sheath that will not allow advancement of the catheter through the peelable sheath. The SmartPort Plastic and Smart Port CT kits may also be packaged with incorrect sized locking collar or port assemblies which may not allow a secure connection of the locking collar to the port stem or not allow the catheter to pull off the port stem once assembled with a smaller locking collar.
Additional information
Details
Angiodynamics, Inc. Also D.B.A. As Navilyst Medical , Inc.
26 Forest Street, Marlborough, Massachusetts, United States, 01752
Get notified