Spacelabs display cables for Xprezzon® or Qube Bedside Monitors
Brand(s)
Last updated
Summary
Product
Spacelabs display cables for Xprezzon® or Qube Bedside Monitors
Issue
Medical devices - Device compatibility
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen | 012-089X-XX (cable) | 94267 |
DVI Cables Used With Qube Compact Monitor | 012-089X-XX (cable) | 91390 |
DVI Cable Used With Bedside Monitor | 012-089X-XX (cable) | 91393 |
Issue
While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Spacelabs Healthcare Inc.
35301 Se Center Street, Snoqualmie, Washington, United States, 98065
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78311
Get notified
Receive emails about new and updated recall and safety alerts.