Health product recall

Spacelabs Healthcare Xhibit® Central Station

Last updated

Summary

Product
Spacelabs Healthcare Xhibit® Central Station
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Spacelabs Healthcare Xhibit® Central Station Not applicable. 96102

Issue

Spacelabs Healthcare has concluded that the use of the previously distributed splitter (part number 010-1975-00) can have an adverse effect on product performance, specifically around audible and visible alarms. Splitters may fail to perform as intended and result in degraded or complete loss of alarm audio and video. Additionally, customer's own use of unqualified splitters may pose the same risks.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Spacelabs Healthcare Inc.

35301 Se Center Street, Snoqualmie, Washington, United States, 98065

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82410

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe