Health product recall

Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

Last updated

Summary

Product
Spring-Wire Guide, Pressure Injectable PICC Kit, Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalogue number

Spring-Wire Guide

More than 10 numbers, contact manufacturer.

PW-18080-HN

Spring-Wire Guide

More than 10 numbers, contact manufacturer.

PW-18080-HS

Spring-Wire Guide

More than 10 numbers, contact manufacturer.

PW-16130-HN

Pressure Injectable PICC Kit

More than 10 numbers, contact manufacturer.

PR-35552-HPHNL

Pressure Injectable PICC Kit

More than 10 numbers, contact manufacturer.

PR-35541-HPHNL

Pressure Injectable PICC Kit

More than 10 numbers, contact manufacturer.

PR-35552-HPHNM

Pressure Injectable PICC Kit

More than 10 numbers, contact manufacturer.

PR-35541-HPHNM

Pressure Injectable PICC Kit

More than 10 numbers, contact manufacturer.

PR-35563-HPHNL

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45552-HPHNL

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45552-HPHNM

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45563-HPHNL

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45541-HPHNM

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45563-HPHNM

Arrow PICC With Arrowg+ard Blue Advance® Technology, No Placement Wire

More than 10 numbers, contact manufacturer.

PR-45541-HPHNL

Issue

Teleflex has initiated a voluntary recall for specific lots of the above-listed products because there is a potential quality issue affecting the supplied hydrophilic coated spring wire guide (SWG) components. Teleflex has confirmed the coating may swell inconsistently when hydrated with saline leading to clumping of the coating and potential for detachment.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Health products - Medical devices - General hospital and personal use
Companies

Arrow International LLC Subsidiary Of Teleflex Incorporated

3015 Carrington Mill Blvd., Morrisville, North Carolina, United States, 27560

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-82973