Health product recall

Stardrive Screwdriver Shaft/T4 50mm/Self-Retaining/Hxc

Brand(s)
Last updated

Summary

Product
Stardrive Screwdriver Shaft/T4 50mm/Self-Retaining/Hxc
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Stardrive Screwdriver Shaft/T4 50mm/Self-Retaining/Hxc

97P6250

03.130.010

Issue

Depuy Synthes is initiating a product removal (recall) for 3 lots of T4 screwdrivers because they were manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3 and 1.5mm screws.

Recall Start Date: May 1, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Synthes Gmbh

Luzernstrasse 21, Zuchwil, Switzerland, 4528

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75524

Get notified

Receive notifications for new and updated recalls and alerts by category.