StatStrip® Glucose And Beta-Ketone Hospital Meter
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
StatStrip® Glucose And Beta-Ketone Hospital Meter |
All firmware versions |
63910 |
Issue
A software issue in StatStrip (Gen 2) Glucose and Beta-Ketone Hospital Meters may retain previous patient information during barcode scanning, potentially assigning results to the wrong patient. Users should verify patient identification before accepting scanned data and follow the correction instructions provided by Nova Biomedical.
Additional information
Details
Nova Biomedical Corporation
200 Prospect St., Waltham, Massachusetts, United States, 02454
Get notified