Health product recall

StatStrip® Glucose And Beta-Ketone Hospital Meter

Last updated

Summary

Product
StatStrip® Glucose And Beta-Ketone Hospital Meter
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

StatStrip® Glucose And Beta-Ketone Hospital Meter

All firmware versions

63910

Issue

A software issue in StatStrip (Gen 2) Glucose and Beta-Ketone Hospital Meters may retain previous patient information during barcode scanning, potentially assigning results to the wrong patient. Users should verify patient identification before accepting scanned data and follow the correction instructions provided by Nova Biomedical.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Nova Biomedical Corporation

200 Prospect St., Waltham, Massachusetts, United States, 02454

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82503

Get notified