Stryker Instruments System 8 and 9 Saws
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
System 8 Reciprocating Saw |
2409624823 |
8206-000-000 |
System 9 Sternum Saw |
Not applicable. |
9207-000-000 |
System 9 Reciprocating Saw |
More than 10 numbers, contact manufacturer. |
9206-000-000 |
Issue
There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade.
Recall Start Date: March 14, 2025
Additional information
Details
Stryker Instruments, A Division Of Stryker Corporation
1941 Stryker Way, Portage, Michigan, United States, 49002
Get notified