Suture/Laceration Tray by Canadian Hospital Specialties Ltd.
Last updated
Summary
Product
Suture/Laceration Tray by Canadian Hospital Specialties Ltd.
Issue
Medical devices - Labelling and packaging
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Suture/Laceration Tray | 208920 | 85-4361 |
Issue
The recall is due to a labeling error affecting the Laceration Tray MRX. The kits that manufactured with lot #208920, have discrepancy in the content and the product label. The original product contains fine teeth forceps; however, the affected lot states Adson forceps without teeth on the label. Customers should check their inventory and destroy them as per their facility process if they are part of the affected lot.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Canadian Hospital Specialties Ltd. Doing Business As CHS
2760 Brighton Road, Oakville, Ontario, Canada, L6H 5T4
Published by
Health Canada
Audience
Industry
Recall class
Type II
Recall date
Identification number
RA-81386
Get notified
Receive emails about new and updated recall and safety alerts.