Health product recall

Sysmex Automated Blood Coagulation Analyzers (2020-09-10)

Starting date:
September 10, 2020
Posting date:
December 18, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74595



Last updated:
2020-12-18

Affected Products

A. CA-540

B. CA-540

C. CA-560

D. CS-20001

E. CS-21001

F. CA-620

G. CA-660

H. CS-5100

I. CS-2500

Reason

The manufacturer of the Sysmex Automated Blood Coagulation Analyzers, Sysmex Corporation, has informed Siemens Healthcare Diagnostics about a potential risk of sample contamination by Emicizumab (trade name Hemlibra®) from a sample containing this drug to the next sample analyzed on one of the Sysmex Automated Blood Coagulation Analyzers.

Affected products

A. CA-540

Lot or serial number

All lots.

Model or catalog number

CA-1500

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


B. CA-540

Lot or serial number

All lots.

Model or catalog number

B4260-540

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


C. CA-560

Lot or serial number

All lots.

Model or catalog number

B4260-560

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


D. CS-20001

Lot or serial number

All lots.

Model or catalog number

10488584

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


E. CS-21001

Lot or serial number

All lots.

Model or catalog number

10488586

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


F. CA-620

Lot or serial number

All lots.

Model or catalog number

10712040

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


G. CA-660

Lot or serial number

All lots.

Model or catalog number

CA-620

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


H. CS-5100

Lot or serial number

All lots.

Model or catalog number

CS-5100

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN


I. CS-2500

Lot or serial number

All lots.

Model or catalog number

CS-2500

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN