Health product recall

Sysmex DI-60™ Main Unit

Last updated

Summary

Product
Sysmex DI-60™ Main Unit
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Sysmex DI-60™ Main Unit All lots. CC286297

Issue

When analyzing a slide with unreadable barcode in a DI-60, there is a risk of incorrect barcode reading and misattribution of results. This risk is higher with certain readers and barcode placements. Interim measures are advised as per notice, and firmware upgrades will be provided to affected units.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies

Sysmex Corporation

4-4-4- Takatsukadai, Kobe, Japan, 651-2271

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78384

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe