Health product recall

Sysmex XN-L Series (2017-12-01)

Starting date:
December 1, 2017
Posting date:
January 8, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65566

Affected Products

Sysmex XN-L Series

Reason

After the change to standard time from Daylight Savings Time, the analyzer's time may not be consistent with local time. The analyzer date may also not reflect the current date. The discrepancy is due to a time difference between the windows operating system and the XN-L program time. After changing the time and date in the analyzer's software and cycling power to the analyzer, the date and time may not be correct when the analyzer restarts. Patient and Quality Control (QC) results are not impacted. However, the analysis date and time displayed, printed and transmitted with the results may not reflect the actual date and time the sample was run.

Affected products

Sysmex XN-L Series

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

XN-450

XN-550