Health product recall

System Control N+P, STA - Coag Control N+P & Unicalibrator - STA (2019-11-19)

Starting date:
November 19, 2019
Posting date:
December 13, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71861



Last updated: 2019-12-13

Affected Products

  1. System Control N+P, STA - Coag Control N+P
  2. Unicalibrator, STA

Reason

Following a customer complaint, Stago has investigated and confirmed that a positive bias exists on the STA - Unicalibrator levels with STA - deficient IX and STA - deficient VIII when compared to the international Standard (IS) for factor VIII or IX. This bias is also observed on several external quality assessment results.

Affected products

A. System Control N+P, STA - Coag Control N+P

Lot or serial number

253145

253416

253655*

254040*

254206

254401

254573

254760

Model or catalog number

678

Companies
Manufacturer

Diagnostica Stago SAS

9, Rue Des Freres Chausson

Asnieres

92600

FRANCE


B. Unicalibrator, STA

Lot or serial number

253180

253964

254721

255108

Model or catalog number

675

Companies
Manufacturer

Diagnostica Stago SAS

9, Rue Des Freres Chausson

Asnieres

92600

FRANCE