Team3 Intrapartum and Team3 Antepartum
Brand(s)
Last updated
Summary
Product
Team3 Intrapartum and Team3 Antepartum
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Team3 Intrapartum | More than 10 numbers, contact manufacturer. | TEAM3I |
Team3 Antepartum | More than 10 numbers, contact manufacturer. | TEAM3A |
Issue
As a follow up to the section 64 report sent on April 3, 2023, it has come to our attention that while the upgrade to version 19.4 software has been successful in the vast majority of cases, we have identified an issue when upgrading some units. The issue occurs when a large amount of data has been retained in the Team3 patient database, resulting in the unit resetting when the database is searched.
Recall start date: May 5, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall