Health product recall

THS™ Neonatal Cart With Powervar Security II Medical UPM

Last updated

Summary

Product
THS™ Neonatal Cart With Powervar Security II Medical UPM
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

THS™ Neonatal Cart With Powervar Security II Medical UPM

5022072R-2330003
5022072R-2330006
5022072R-2330005
5022072R-2330001
5022072R-2330004
5022072R-2330002

50220-72R

Issue

Dräger has been made aware of an incident involving the UPS battery on the Technimount THS™ Neonatal Cart switching off during patient transfer with the Dräger incubator Babyleo® TN500 (Health Canada file no. 230304). There were no injuries or impact on patient care resulting from this incident, however Dräger Medical Canada Inc., the  distributor of the cart has deemed necessary to initiate a recall to advise users to adhere to the battery IFU to prevent a similar failure.

Recall start date: October 1, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Draegerwerk AG & Co. KGaA

Moislinger Allee 53-55, Lubeck, Germany, 23542

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76379

Get notified

Receive notifications for new and updated recalls and alerts by category.