Health product recall

Trackmaster Treadmills

Brand(s)
Last updated

Summary

Product
Trackmaster Treadmills
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

Trackmaster Treadmills

Contact manufacturer.

TMX428 220
TMX428 110

Issue

This involves the motor/drive pcb combination of the GE Healthcare T2100-ST series and the Trackmaster series treadmills.  The motor/drive pcb configuration has inherent noise between the encoder and the drive pcb which may cause the drive pcb to misread the orientation of the motor when commanded to start movement.  This belt may accelerate forwards or backwards without notice.

Recall Start Date: May 24, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies

Full Vision Inc

3017 Full Vision Dr., Newton, Kansas, United States, 67114

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73913

Get notified

Receive notifications for new and updated recalls and alerts by category.