UHI-4 High Flow Insufflation System - Insufflator
Brand(s)
Last updated
Summary
Product
UHI-4 High Flow Insufflation System - Insufflator
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| UHI-4 High Flow Insufflation System - Insufflator | More than 10 numbers, contact manufacturer. | UHI-4 |
Issue
Olympus identified an increase in UHI-4 front-panel 'blackout' complaints, where the front-panel display turns off unexpectedly, and air delivery (insufflation) ceases. An Olympus field service representative will contact affected users to schedule a mutually convenient time to perform the offset adjustment.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Olympus Medical Systems Corp.
2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82247