UniCel DxI 600 & 800 Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
UniCel DxI 600 Access Immunoassay Analyzer | 902386 | A30260 |
UniCel DxI 600 Access Immunoassay Analyzer | 902624 | A30260 |
UniCel DxI 600 Access Immunoassay Analyzer | 902385 | A30260 |
UniCel DxI 600 Access Immunoassay Analyzer | 902622 | A30260 |
UniCel DxI 600 Access Immunoassay Analyzer | 902475 | A30260 |
DxI 600 Access Immunoassay W/Spot B | More than 10 numbers, contact manufacturer. | A71460 |
UniCel DxI 800 Access Immunoassay Analyzer | 609576 | 973100 |
Issue
Beckman identified an issue on UniCel DxI 600 & 800 Systems with linear slide pnp gantries running software versions 5.7 and up. During installs or upgrades, the Beckman identified an issue on UniCel DxI 600 & 800 Systems with linear slide pnp gantries running software versions 5.7 and up. During installs or upgrades, the pnp xmotor current incorrectly resets to 1.5a (instead of the required 2.0a) because the software overwrites the factory setting, causing increased pnp motion errors. current incorrectly resets to 1.5a (instead of the required 2.0a) because the software overwrites the factory setting, causing increased pnp motion errors.
Recall start date: June 18, 2025