Health product recall

UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB (2019-04-13)

Starting date:
April 13, 2019
Posting date:
May 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69834



Last updated: 2019-05-03

Affected Products

A. UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB

B. Access System - Intrinsic Factor AB

C. Intrinsic Factor AB Assay

Reason

Beckman Coulter has determined for the Access Intrinsic Factor Antibody (IFab) assay that in rare cases a "negative" result interpretation, rather than "equivocal", is assigned when the result is between 1.20 and 1.2000000476 AU/mL.

Affected products

A. UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB

Lot or serial number

All lots.

Model or catalog number

387992