Health product recall

Universal Joint Screwdriver

Last updated

Summary

Product
Universal Joint Screwdriver
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalog number

Universal Joint Screwdriver

1000558234
1000555951
1000549586
1000530579
1000530603
1000532954
1000553021

6003-100-110

Issue

The tip component of the Universal Joint Screwdriver may break off during use. The devices utilize a clamping tool that delivered too much force during manufacturing, resulting in product deformation.

Recall Start Date: February 27, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies

Stryker Leibinger Gmbh & Co. Kg

Botzinger Strasse 41, Freiburg, Germany, 79111

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-77085

Get notified

Receive notifications for new and updated recalls and alerts by category.