Health product recall

Vantage Galan 3T

Last updated

Summary

Product
Vantage Galan 3T
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Vantage Galan 3T Y5C1922001 MRT-3020/5Y

Issue

It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Canon Medical Systems Corporation

1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78458