Vantage Gastroscopes and Colonoscopes
Brand(s)
Last updated
Summary
Product
Vantage Gastroscopes and Colonoscopes
Issue
Medical devices - Performance
What to do
Contact the importer Vantage Endoscopy.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Video Gastroscope Therapeutic Type; Working Length 1100mm | 8G411K033 | EG-760CT |
3D Navigation Colonoscope | 1C754K057 | EC-760S-G/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 2C731K155 | EC-760ZP-V/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | KG402K232 | EG-760R |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 8C729K097 | EC-760R-V/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | JC729K932 | EC-760R-V/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 7C729K100 | EC-760R-V/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 7G402K360 | EG-760R |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | JG402K452 | EG-760R |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 5C729K533 | EC-760R-V/L |
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 5C729K534 | EC-760R-V/L |
720 Adult Colonoscope | 1C742K226 | EC-720R/L |
Dual Channel Endoscope | 2G417K087 | EI-740D/S |
Issue
The devices were recently repaired at Vantage's service centre and returned to the customers. It came to their attention that the irrigation water pressure of these device(s) may exceed the specified range, potentially resulting in stronger-than-expected pressure during a procedure.
Recall start date: May 5, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General and plastic surgery
Companies
Fujifilm Corporation
26-30 Nishiazabu, Minato-Ku, Tokyo-To, Japan, 106-8620
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77549