Health product recall

Vantage Gastroscopes and Colonoscopes

Last updated

Summary

Product
Vantage Gastroscopes and Colonoscopes
Issue
Medical devices - Performance
What to do

Contact the importer Vantage Endoscopy.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Video Gastroscope Therapeutic Type; Working Length 1100mm

8G411K033

EG-760CT

3D Navigation Colonoscope

1C754K057

EC-760S-G/L

Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8

2C731K155
KG402K232
8C729K097
JC729K932
7C729K100
7G402K360
JG402K452
5C729K533
5C729K534

EC-760ZP-V/L
EG-760R
EC-760R-V/L

720 Adult Colonoscope

1C742K226

EC-720R/L

Issue

The devices were recently repaired at Vantage's service centre and returned to the customers. It came to their attention that the irrigation water pressure of these device(s) may exceed the specified range, potentially resulting in stronger-than-expected pressure during a procedure.

Recall start date: May 5, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General and plastic surgery
Companies

Fujifilm Corporation

26-30 Nishiazabu, Minato-Ku, Tokyo-To, Japan, 106-8620

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77549

Get notified

Receive notifications for new and updated recalls and alerts by category.