Health product recall

Vantage Gastroscopes and Colonoscopes

Last updated

Summary

Product
Vantage Gastroscopes and Colonoscopes
Issue
Medical devices - Performance
What to do

Contact the importer Vantage Endoscopy.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Video Gastroscope Therapeutic Type; Working Length 1100mm 8G411K033 EG-760CT
3D Navigation Colonoscope 1C754K057 EC-760S-G/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 2C731K155 EC-760ZP-V/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 KG402K232 EG-760R
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 8C729K097 EC-760R-V/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 JC729K932 EC-760R-V/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 7C729K100 EC-760R-V/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 7G402K360 EG-760R
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 JG402K452 EG-760R
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 5C729K533 EC-760R-V/L
Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 5C729K534 EC-760R-V/L
720 Adult Colonoscope 1C742K226 EC-720R/L
Dual Channel Endoscope 2G417K087 EI-740D/S

Issue

The devices were recently repaired at Vantage's service centre and returned to the customers. It came to their attention that the irrigation water pressure of these device(s) may exceed the specified range, potentially resulting in stronger-than-expected pressure during a procedure.

Recall start date: May 5, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General and plastic surgery
Companies

Fujifilm Corporation

26-30 Nishiazabu, Minato-Ku, Tokyo-To, Japan, 106-8620

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77549

Get notified

Receive notifications for new and updated recalls and alerts by category.