Vantage Titan HSR (2018-05-09)
- Starting date:
- May 9, 2018
- Posting date:
- June 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66944
Affected products
Vantage Titan HSR
Reason
When an operator uses the respiratory gating with steady mode on in spin echo sequence scan, the SAR value displayed on the console may be lower than the actual SAR value.
Affected products
Vantage Titan HSR
Lot or serial number
Y5A10Z2001
Model or catalog number
MRT-1504_/Y5
Companies
- Manufacturer
-
Toshiba Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi-ken
324-8550
JAPAN