Health product recall

VascuPuncture™ PICC High Performance Guidewire

Last updated

Summary

Product
VascuPuncture™ PICC High Performance Guidewire
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Hospitals
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

VascuPuncture™ PICC High Performance Guidewire

More than 10 numbers, contact manufacturer.

18070280NW-05MHP

VascuPuncture™ PICC High Performance Guidewire

More than 10 numbers, contact manufacturer.

18145280NW-05MHP

Issue

Hydrophilic coated guidewire may swell inconsistently when exposed to saline, causing coating clumping and potential particle detachment. Released particles could enter the bloodstream and potentially cause embolic events.

Users should immediately discontinue use of affected devices, quarantine inventory, and follow recall instructions for return.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Heraeus Medical Components, LLC

5030 Centerville Road, Saint Paul, Minnesota, United States, 55127

Published by
Health Canada
Audience
Hospitals
Industry
Recall class
Type I
Recall date
Identification number
RA-82976

Get notified