Vivo 45 LS
Brand(s)
Last updated
Summary
Product
Vivo 45 LS
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Vivo 45 LS | S/N: 1R070KF to 1R100F9 | 230000 |
Issue
In devices with the affected firmware versions, treatment in volume-controlled ventilation modes (VCV, VCV-SIMV and VCV-MPV) can, under specific trigger conditions, result in delivered tidal volume being reduced significantly below the set value.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Breas Medical AB
Foretagsvagen 1, Molnlycke, Sweden, 43533
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82020
Get notified