Volista Surgical Lights (2018-05-24)
- Starting date:
- May 24, 2018
- Posting date:
- June 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66950
Affected products
Volista Surgical Lights
Reason
It was established that, with some of the manufactured units, a screw has been forgotten on the bracket of the cupola on some devices in production or that the screws have loosened over the course of life. As a consequence the light head shaft could break and the cupola could detach itself when manipulated and be held only by its cables. It appears such event is more likely to happen mainly during maintenance, cleaning and surgery preparation.
Affected products
Volista Surgical Lights
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
ARD568811903
ARD568811912
ARD568811952
ARD568811961
ARD568811963
ARD568812912
ARD568812952
ARD568812963
ARD568821960
ARD568821961
ARD568821963
ARD568822963
Companies
- Manufacturer
-
Maquet S.A.
Avenue de la Pomme De Pin, Parc De Limere
Ardon, Orleans
45074
FRANCE