Health product recall

Wash/Waste Cartridge For RAPIDPoint® 500 And RAPIDPoint® 500e

Last updated

Summary

Product
Wash/Waste Cartridge For RAPIDPoint® 500 And RAPIDPoint® 500e
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Wash/Waste Cartridge For RAPIDPoint® 500 And RAPIDPoint® 500e WW/08925 10329097 (129672)

Issue

Siemens Healthcare Diagnostics has confirmed, through complaint investigation, that some RAPIDPoint® 500 Systems wash/waste cartridges from lot WW/08925 may contain an incorrect wash pack.

This issue may affect the test results of the following measured analytes: ionized calcium, potassium, pH, chloride, pCO2, glucose and lactate.

Recall start date: July 2, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave, Tarrytown, New York, United States, 10591

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77734

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe