Welch Allyn Connex Vital Signs Monitor and Electrocardiograph
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Vital Signs Monitor (CVSM) |
All lots. |
Contact manufacturer. |
Welch Allyn Cp150 Electrocardiograph |
All lots. |
Contact manufacturer. |
Welch Allyn Connex Spot Monitor Accessory Power Management Stand (Apm) - Connex Spot Monitor |
All lots. |
BATT99 |
Issue
Baxter Corporation is issuing urgent medical device correction for devices which experienced battery-related fires due to use of third-party, after-market batteries (not approved by Baxter). Baxter will update IFUs and service manuals to include instructions to only use Baxter approved battery for replacement and highlight risks associated with using unapproved battery.
Recall start date: March 19, 2025
Additional information
Details
Welch Allyn, Inc. |
4341 State Street Road, Skaneateles Falls, New York, United States, 13153 |