XERF Effector 60
Issue
Medical devices - Performance
What to do
Customers will be contacted with instructions to return and replace affected units. |
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
XERF Effector 60 | More than 10 numbers, contact manufacturer. | 1101210000 |
XERF Effector 60 | More than 10 numbers, contact manufacturer. | 1101210100 |
XERF Effector 60 | More than 10 numbers, contact manufacturer. | 1101210200 |
Issue
Lutronic has identified a manufacturing issue that results in XERF Effector 60 tips with micro-scratches on the electrode surface. There is a remote possibility that use of such tips will lead to tip failure and potential for patient injury.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Lutronic Corporation |
Lutronic Center, Goyang-Si, Gyeonggi-Do, South Korea, 412-220 |
Published by
Health Canada
Audience
General public
Recall class
Type II
Recall date
Identification number
RA-78070