Health product recall

XP003/20/P Mammomat Revelation AMC Box: Grey Images (2020-04-08)

Starting date:
April 8, 2020
Posting date:
May 8, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72993



Last updated: 2020-05-08

Affected Products

XP003/20/P Mammomat Revelation AMC Box: Grey Images

Reason

Siemens has identified some issues with the AMC box on the multitom rax unit. The system sporadically produces grey images that are being generated via the one-point technique. The issue is caused by an AMC box (automatic exposure control multic converter). A new revision level of the AMC box is being implemented. It contains a change of the power supply and implementation of an additional pull-down resistor. Siemens is releasing this update instruction to provide new amc box to resolve the issue of sporadically generated grey images.

Affected products

XP003/20/P Mammomat Revelation AMC Box: Grey Images

Lot or serial number

11017

Model or catalog number

10860700

Companies
Manufacturer

Siemens Healthcare GMBH

Henkestr. 127

Erlangen

91052

GERMANY