Ysio X.pree (2021-03-12)
- Starting date:
- March 12, 2021
- Posting date:
- April 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75277
Last updated:
2021-04-01
Affected Products
Ysio X.pree
Reason
Siemens has received a complaint about an image assigned to a different patient during image recovery process. In the unlikely event of an interrupted connection to a max detector during an initiated image recovery, the previously acquired image (possibly belonging to the previous patient) might be displayed to the user instead of the currently acquired image.
Affected products
Ysio X.pree
Lot or serial number
- VA10C
Model or catalog number
- 11107464