Enlite & Guardian Sensor (2021-03-08)
- Date de début :
- 8 mars 2021
- Date d’affichage :
- 8 avril 2021
- Type de communication :
- Rappel d'instruments médicaux
- Sous-catégorie :
- Instruments médicaux
- Classification du risque :
- Type II
- Source :
- Santé Canada
- Problème :
- Instruments médicaux
- Public :
- Hôpitaux, Professionnels de la santé, Grand public
- Numéro d’identification :
- RA-75331
Dernière mise à jour:
2021-04-08
Produits Touchés
A. Enlite Sensor
B. GUARDIAN SENSOR (3)
Raison
Medtronic is notifying all customers who may be using an Enhanced Enlite Sensor or a Guardian Sensor (3) in a Continuous Glucose Monitoring (CGM) system or in a sensor-enabled pump system that, if they are actively taking a medication that lists Hydroxyurea as its main or one of its main ingredients, their sensor could present inaccurately elevated Sensor Glucose (SG) readings. Hydroxyurea use results in higher SG readings compared to Blood Glucose (BG) readings and may result in the following: Hypoglycemia caused by over-administration of insulin. Inaccurate graphs or missed alarms and alerts. Delay to or loss of sensor-enabled insulin suspension (sensor-enabled insulin pump use only). SG readings in CareLink reports being substantially higher than BG readings.
Produits touchés
A. Enlite Sensor
Numéro de lot ou de série
Plus de 10 numéros, contactez le fabricant.
Numéro de modèle ou de catalogue
MMT-7008A
Entreprises
- Fabricant
-
Medtronic Minimed
18000 Devonshire St.
Northridge
91325
Californie
ÉTATS-UNIS
B. GUARDIAN SENSOR (3)
Numéro de lot ou de série
Plus de 10 numéros, contactez le fabricant.
Numéro de modèle ou de catalogue
MMT-7020C4
MMT-7020D4
Entreprises
- Fabricant
-
Medtronic Minimed
18000 Devonshire St.
Northridge
91325
Californie
ÉTATS-UNIS