Health product recall

ACCOLADE™ and VISIONIST™ CRT-P Pacemakers

Last updated

Summary

Product
ACCOLADE™ and VISIONIST™ CRT-P Pacemakers
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
ACCOLADE™ Pacemaker More than 10 numbers, contact manufacturer. L301
ACCOLADE™ Pacemaker More than 10 numbers, contact manufacturer. L321
ACCOLADE™ Pacemaker More than 10 numbers, contact manufacturer. L300
ACCOLADE™ MRI Pacemaker More than 10 numbers, contact manufacturer. L310
ACCOLADE™ MRI Pacemaker More than 10 numbers, contact manufacturer. L311
ACCOLADE™ MRI Pacemaker More than 10 numbers, contact manufacturer. L331
VISIONIST™ CRT-P More than 10 numbers, contact manufacturer. U225
VISIONIST™ CRT-P More than 10 numbers, contact manufacturer. U226
VISIONIST™ X4 Crt-P More than 10 numbers, contact manufacturer. U228
ESSENTIO Pacemaker More than 10 numbers, contact manufacturer. L100
ESSENTIO Pacemaker More than 10 numbers, contact manufacturer. L101
ESSENTIO Pacemaker More than 10 numbers, contact manufacturer. L121
ESSENTIO MRI Pacemaker More than 10 numbers, contact manufacturer. L110
ESSENTIO MRI Pacemaker More than 10 numbers, contact manufacturer. L111
ESSENTIO MRI Pacemaker More than 10 numbers, contact manufacturer. L131
PROPONENT Pacemaker More than 10 numbers, contact manufacturer. L200
PROPONENT Pacemaker More than 10 numbers, contact manufacturer. L201
PROPONENT Pacemaker More than 10 numbers, contact manufacturer. L209
PROPONENT Pacemaker More than 10 numbers, contact manufacturer. L221
PROPONENT MRI Pacemaker More than 10 numbers, contact manufacturer. L210
PROPONENT MRI Pacemaker More than 10 numbers, contact manufacturer. L211
PROPONENT MRI Pacemaker More than 10 numbers, contact manufacturer. L231
ALTRUA 2 Pacemaker More than 10 numbers, contact manufacturer. S701
ALTRUA 2 Pacemaker More than 10 numbers, contact manufacturer. S702
ALTRUA 2 Pacemaker More than 10 numbers, contact manufacturer. S722
VALITUDE CRT-P More than 10 numbers, contact manufacturer. U125
VALITUDE X4 Crt-P More than 10 numbers, contact manufacturer. U128

Issue

Boston Scientific is initiating an advisory to provide information about a subset of pacemakers from the ACCOLADE™ family and VISIONIST™ CRT-Ps with an increased potential to initiate safety mode during telemetry or, in rare instances, other normal, higher-power operations due to latent, high battery impedance when the device reaches approximately four (4) years or less of remaining battery longevity.

Recall start date: December 12, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Boston Scientific Corporation

4100 Hamline Avenue North, St. Paul, Minnesota, United States, 55112-5798

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-76710

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe