ADVIA Chemistry System - Urea Nitrogen (UN) Assay (2018-04-05)
- Starting date:
- April 5, 2018
- Posting date:
- May 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66632
Affected products
ADVIA Chemistry System - Urea Nitrogen (UN) Assay
Reason
Siemens Healthcare Diagnostics has confirmed that the Urea Nitrogen Reagent Kit Lots 408143 and 418290 for use on ADVIA 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of calibration failure. Results cannot be generated by the system when calibration fails. If calibration passes, there is a potential for falsely elevated patient and quality control results. Internal testing has also shown that results have the potential to be falsely increased by approximately 10 mg/dl (3.57 mmol/l) across the analytical range. Quality controls may not always detect the elevated results.
Affected products
ADVIA Chemistry System - Urea Nitrogen (UN) Assay
Lot or serial number
408143
Model or catalog number
03040257
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES