ARCHITECT BNP Controls (2019-05-29)
- Starting date:
- May 29, 2019
- Posting date:
- June 14, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70199
Last updated: 2019-06-14
Affected Products
ARCHITECT BNP Controls
Reason
Falsely elevated patient and/or control results may be obtained when using ARCHITECT BNP calibrators and controls. Preliminary investigation has determined that all current in-date BNP calibrator and control lots are impacted by an age dependent shift in performance. Use of impacted calibrator lots may result in falsely elevated patient results. Use of impacted control lots may result in invalid patient results due to controls out of range.
Affected products
ARCHITECT BNP Controls
Lot or serial number
- 44K78918
- 44K79318
- 44K80818
- 44K80918
- 44K82118
- 44K82618
- 44K82718
- 44K85119
Model or catalog number
- 8K28-02
- 8K28-11
Companies
- Manufacturer
-
Abbott Laboratories Diagnostics Division
100 ABBOTT PARK ROAD
ABBOTT PARK
60064
Illinois
UNITED STATES