Arjo Tenor Mobile Lift
Brand(s)
Last updated
Summary
Product
Arjo Tenor Mobile Lift
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Arjo Tenor Mobile Lift | All lots. | KHA1000-US |
| Arjo Tenor Mobile Lift | All lots. | TEN.107 |
| Arjo Tenor Mobile Lift | All lots. | KHA1010-US |
Issue
A supplier‑related brake component defect in the Tenor lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load. All affected units must be removed from use immediately and remain out of service until actuator replacement is completed by authorized Arjo service personnel.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Arjohuntleigh Ab
Hans Michelsensgatan 10, Malmo, Sweden, 21120
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81735
Get notified