Health product recall

Arjo Tenor Mobile Lift

Brand(s)
Last updated

Summary

Product
Arjo Tenor Mobile Lift
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Arjo Tenor Mobile Lift All lots. KHA1000-US
Arjo Tenor Mobile Lift All lots. TEN.107
Arjo Tenor Mobile Lift All lots. KHA1010-US

Issue

A supplier‑related brake component defect in the Tenor lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load. All affected units must be removed from use immediately and remain out of service until actuator replacement is completed by authorized Arjo service personnel.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Arjohuntleigh Ab

Hans Michelsensgatan 10, Malmo, Sweden, 21120

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81735

Get notified

Receive emails about new and updated recall and safety alerts.