Health product recall

Automated Impella Controller

Brand(s)
Last updated

Summary

Product
Automated Impella Controller
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Automated Impella Controller Software version 8.2.3 0042-0040-CA
Automated Impella Controller Software version 8.2.3 0042-0000-CA

Issue

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. 
Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Abiomed Inc.

22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-81652