BD BBL Gram Stain Kit & Crystal Violet (2020-04-20)
- Starting date:
- April 20, 2020
- Posting date:
- May 8, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73025
Last updated: 2020-05-22
Affected Products
- BD BBL Gram Stain Kit
- BD BBL Gram Crystal Violet
Reason
Confirmed customer reports that the products may exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci. Use of products with increased levels of artifacts could result in a false-positive gram stain. Gram stain results must be confirmed with additional procedures, but if these protocols are not followed, a false-positive could lead to administration of unnecessary antibiotics or surgical procedures.
Affected products
A. BD BBL Gram Stain Kit
Lot or serial number
- 7255
- 7261
- 14322
- 411114
Model or catalog number
- 212539