Health product recall

BD BBL Sensi Disc Ampicillin 2 Μg (AM-2)

Last updated

Summary

Product
BD BBL Sensi Disc Ampicillin 2 Μg (AM-2)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

BD BBL Sensi Disc Ampicillin 2 Μg (AM-2)

3234190
3184064
2339360
3010977

231263

Issue

BD identified through potency test, as part of special stability test requests to monitor Ampicillin AM-2, unsatisfactory results of 65% at 18 months. The potency acceptance criteria is 80-140% and product shelf life is 24 months. A retest with control batch and extended sampling confirmed the low potency results.

Recall Start Date: November 6, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies

Becton Dickinson And Company (Bd)

7 Loveton Circle, Sparks, Maryland, United States, 21152

Published by
Health Canada
Audience
Industry
Recall class
Type II
Identification number
RA-76548

Get notified

Receive notifications for new and updated recalls and alerts by category.