BD FACS™ 7-Color Setup Beads
Last updated
Summary
Product
BD FACS™ 7-Color Setup Beads
Issue
Medical devices - Labelling and packaging
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
BD FACS™ 7-Color Setup Beads | 85612 85613 | 335775 |
Issue
Based on customer complaints, BD has identified errors in the Spectral Overlap Factors (SOF) table provided for specific lots of BD FACS™ 7-color setup beads as listed in the table above. This issue may affect the accuracy of flow cytometry data due to incorrect flow cytometry setup compensation values for multicolor reagents for which proper fluorescence compensation is required prior to analyzing samples.
Recall start date: June 3, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Becton, Dickinson And Company Bd Biosciences |
155 North Mccarthy Boulevard, Milpitas, California, United States, 95035 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77584
Get notified
Receive notifications for new and updated recalls and alerts by category.