BD Texium™ Components
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
BD Texium™ Closed Male Luer With Female Cap |
More than 10 numbers, contact manufacturer. |
10012241-0500 |
BD 20 Drop Secondary Set And BD 31-In Sec Set 20dp |
More than 10 numbers, contact manufacturer. |
10013364T |
Alaris™ Pump Infusion Set SmartSite Bag Access Non-Vented Bonded Texiumtm With Priming Cap Back Check Valve 2 Smartsitetm Ysites |
More than 10 numbers, contact manufacturer. |
10013361T |
Issue
Through complaint investigations, BD has identified that specific lots of standalone Texium™ components, and IV infusion sets and syringes containing the Texium™ closed-male luer component (as listed in attachment 1) have the potential to leak during use at the connection interface between the Texium™ component and the SmartSite™ needle-free connector. BD has confirmed that this issue is limited to the affected products listed in attachment 1.
Recall start date: May 12, 2025
Additional information
Details