Health product recall

Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay

Last updated

Summary

Product
Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue
Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay More than 10 numbers, contact manufacturer. OSR6196
Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay More than 10 numbers, contact manufacturer. OSR6296

Issue

LDL cholesterol OSR6x96 does not meet the triglyceride interference claim per the BAOSR6x96 v10 IFU which states "no significant interference up to 1,500 mg/dl triglyceride" where the criteria for no significant interference is recovery within 10% of the initial value as outlined in the IFU.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81343

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe