BiCOAG Hemostasis Probe
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
BiCOAG Hemostasis Probe - Fixed Pin Connector |
More than 10 numbers, contact manufacturer. |
CD-B622LA |
BiCOAG Hemostasis Probe - Fixed Pin Connector |
More than 10 numbers, contact manufacturer. |
CD-B612LA |
BiCOAG Hemostasis Probe - Coaxial Connector |
More than 10 numbers, contact manufacturer. |
CD-B610LA |
BiCOAG Hemostasis Probe - Coaxial Connector |
More than 10 numbers, contact manufacturer. |
CD-B620LA |
Issue
Olympus determined that the affected products may have been manufactured without the designated amount of adhesive to the rear of the device’s ceramic tip, which may result in detachment of the ceramic tip. Customers should immediately discontinue using the BiCOAG Hemostasis Probe products and contact Olympus to return the product.
Additional information
Details
Gyrus Acmi, Inc
9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445
Get notified